CVE-2017-9657
Philips IntelliVue MX40 802.11 Timing-Based Authentication Bypass
Description
Under specific 802.11 network conditions, a partial re-association of the Philips IntelliVue MX40 Version B.06.18 WLAN monitor to the central monitoring station is possible. In this state, the central monitoring station can indicate the MX40 is not connected or associated to the central monitor, and thus should be operating in local monitoring mode (local audio-on, screen-on), but the MX40 WLAN itself can instead still be operating in telemetry mode (local audio-off, screen-off). If a patient experiences an alarm event and clinical staff expects the MX40 to provide local alarming when it is not available from the local device, a delay of treatment can occur. CVSS v3 base score: 6.5, CVSS vector string: AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H. Philips has released software update, Version B.06.18, to fix the improper cleanup on thrown exception vulnerability, and implement mitigations to reduce the risk associated with the improper handling of exceptional conditions vulnerability. The software update implements messaging and alarming on the MX40 and at the central monitoring station, when the MX40 disconnects from the access point.
INFO
Published Date :
April 30, 2018, 3:29 p.m.
Last Modified :
Nov. 21, 2024, 3:36 a.m.
Source :
[email protected]
Remotely Exploitable :
No
Impact Score :
3.6
Exploitability Score :
2.8
Affected Products
The following products are affected by CVE-2017-9657
vulnerability.
Even if cvefeed.io
is aware of the exact versions of the
products
that
are
affected, the information is not represented in the table below.
References to Advisories, Solutions, and Tools
Here, you will find a curated list of external links that provide in-depth
information, practical solutions, and valuable tools related to
CVE-2017-9657
.
URL | Resource |
---|---|
http://www.securityfocus.com/bid/100813 | Third Party Advisory VDB Entry |
https://ics-cert.us-cert.gov/advisories/ICSMA-17-255-01 | Third Party Advisory US Government Resource |
https://www.usa.philips.com/healthcare/about/customer-support/product-security | Vendor Advisory |
http://www.securityfocus.com/bid/100813 | Third Party Advisory VDB Entry |
https://ics-cert.us-cert.gov/advisories/ICSMA-17-255-01 | Third Party Advisory US Government Resource |
https://www.usa.philips.com/healthcare/about/customer-support/product-security | Vendor Advisory |
We scan GitHub repositories to detect new proof-of-concept exploits. Following list is a collection of public exploits and proof-of-concepts, which have been published on GitHub (sorted by the most recently updated).
Results are limited to the first 15 repositories due to potential performance issues.
The following list is the news that have been mention
CVE-2017-9657
vulnerability anywhere in the article.
The following table lists the changes that have been made to the
CVE-2017-9657
vulnerability over time.
Vulnerability history details can be useful for understanding the evolution of a vulnerability, and for identifying the most recent changes that may impact the vulnerability's severity, exploitability, or other characteristics.
-
CVE Modified by af854a3a-2127-422b-91ae-364da2661108
Nov. 21, 2024
Action Type Old Value New Value Added Reference http://www.securityfocus.com/bid/100813 Added Reference https://ics-cert.us-cert.gov/advisories/ICSMA-17-255-01 Added Reference https://www.usa.philips.com/healthcare/about/customer-support/product-security -
CVE Modified by [email protected]
May. 14, 2024
Action Type Old Value New Value -
CVE Modified by [email protected]
Oct. 09, 2019
Action Type Old Value New Value Added CWE ICS-CERT CWE-460 -
CWE Remap by [email protected]
Oct. 03, 2019
Action Type Old Value New Value Changed CWE CWE-19 CWE-19 CWE-755 CWE-755 -
Initial Analysis by [email protected]
Jun. 12, 2018
Action Type Old Value New Value Added CVSS V2 (AV:A/AC:L/Au:N/C:N/I:N/A:P) Added CVSS V3 AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H Changed Reference Type https://www.usa.philips.com/healthcare/about/customer-support/product-security No Types Assigned https://www.usa.philips.com/healthcare/about/customer-support/product-security Vendor Advisory Changed Reference Type https://ics-cert.us-cert.gov/advisories/ICSMA-17-255-01 No Types Assigned https://ics-cert.us-cert.gov/advisories/ICSMA-17-255-01 Third Party Advisory, US Government Resource Changed Reference Type http://www.securityfocus.com/bid/100813 No Types Assigned http://www.securityfocus.com/bid/100813 Third Party Advisory, VDB Entry Added CWE CWE-19 Added CPE Configuration AND OR *cpe:2.3:o:philips:intellivue_mx40_firmware:*:*:*:*:*:*:*:* versions up to (excluding) b.06.18 OR cpe:2.3:h:philips:intellivue_mx40:-:*:*:*:*:*:*:* -
CVE Modified by [email protected]
May. 02, 2018
Action Type Old Value New Value Added Reference http://www.securityfocus.com/bid/100813 [No Types Assigned]
CWE - Common Weakness Enumeration
While CVE identifies
specific instances of vulnerabilities, CWE categorizes the common flaws or
weaknesses that can lead to vulnerabilities. CVE-2017-9657
is
associated with the following CWEs:
Common Attack Pattern Enumeration and Classification (CAPEC)
Common Attack Pattern Enumeration and Classification
(CAPEC)
stores attack patterns, which are descriptions of the common attributes and
approaches employed by adversaries to exploit the CVE-2017-9657
weaknesses.
Exploit Prediction
EPSS is a daily estimate of the probability of exploitation activity being observed over the next 30 days.
0.12 }} -0.00%
score
0.46013
percentile