6.5
MEDIUM
CVE-2017-9657
Philips IntelliVue MX40 802.11 Timing-Based Authentication Bypass
Description

Under specific 802.11 network conditions, a partial re-association of the Philips IntelliVue MX40 Version B.06.18 WLAN monitor to the central monitoring station is possible. In this state, the central monitoring station can indicate the MX40 is not connected or associated to the central monitor, and thus should be operating in local monitoring mode (local audio-on, screen-on), but the MX40 WLAN itself can instead still be operating in telemetry mode (local audio-off, screen-off). If a patient experiences an alarm event and clinical staff expects the MX40 to provide local alarming when it is not available from the local device, a delay of treatment can occur. CVSS v3 base score: 6.5, CVSS vector string: AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H. Philips has released software update, Version B.06.18, to fix the improper cleanup on thrown exception vulnerability, and implement mitigations to reduce the risk associated with the improper handling of exceptional conditions vulnerability. The software update implements messaging and alarming on the MX40 and at the central monitoring station, when the MX40 disconnects from the access point.

INFO

Published Date :

April 30, 2018, 3:29 p.m.

Last Modified :

Oct. 9, 2019, 11:30 p.m.

Remotely Exploitable :

No

Impact Score :

3.6

Exploitability Score :

2.8
Affected Products

The following products are affected by CVE-2017-9657 vulnerability. Even if cvefeed.io is aware of the exact versions of the products that are affected, the information is not represented in the table below.

ID Vendor Product Action
1 Philips intellivue_mx40_firmware
References to Advisories, Solutions, and Tools

Here, you will find a curated list of external links that provide in-depth information, practical solutions, and valuable tools related to CVE-2017-9657.

URL Resource
http://www.securityfocus.com/bid/100813 VDB Entry Third Party Advisory
https://ics-cert.us-cert.gov/advisories/ICSMA-17-255-01 Third Party Advisory US Government Resource
https://www.usa.philips.com/healthcare/about/customer-support/product-security Vendor Advisory

We scan GitHub repositories to detect new proof-of-concept exploits. Following list is a collection of public exploits and proof-of-concepts, which have been published on GitHub (sorted by the most recently updated).

Results are limited to the first 15 repositories due to potential performance issues.

The following list is the news that have been mention CVE-2017-9657 vulnerability anywhere in the article.

The following table lists the changes that have been made to the CVE-2017-9657 vulnerability over time.

Vulnerability history details can be useful for understanding the evolution of a vulnerability, and for identifying the most recent changes that may impact the vulnerability's severity, exploitability, or other characteristics.

  • CVE Modified by [email protected]

    May. 14, 2024

    Action Type Old Value New Value
  • CVE Modified by [email protected]

    Oct. 09, 2019

    Action Type Old Value New Value
    Added CWE ICS-CERT CWE-460
  • CWE Remap by [email protected]

    Oct. 03, 2019

    Action Type Old Value New Value
    Changed CWE CWE-19 CWE-19 CWE-755 CWE-755
  • Initial Analysis by [email protected]

    Jun. 12, 2018

    Action Type Old Value New Value
    Added CVSS V2 (AV:A/AC:L/Au:N/C:N/I:N/A:P)
    Added CVSS V3 AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H
    Changed Reference Type https://www.usa.philips.com/healthcare/about/customer-support/product-security No Types Assigned https://www.usa.philips.com/healthcare/about/customer-support/product-security Vendor Advisory
    Changed Reference Type https://ics-cert.us-cert.gov/advisories/ICSMA-17-255-01 No Types Assigned https://ics-cert.us-cert.gov/advisories/ICSMA-17-255-01 Third Party Advisory, US Government Resource
    Changed Reference Type http://www.securityfocus.com/bid/100813 No Types Assigned http://www.securityfocus.com/bid/100813 Third Party Advisory, VDB Entry
    Added CWE CWE-19
    Added CPE Configuration AND OR *cpe:2.3:o:philips:intellivue_mx40_firmware:*:*:*:*:*:*:*:* versions up to (excluding) b.06.18 OR cpe:2.3:h:philips:intellivue_mx40:-:*:*:*:*:*:*:*
  • CVE Modified by [email protected]

    May. 02, 2018

    Action Type Old Value New Value
    Added Reference http://www.securityfocus.com/bid/100813 [No Types Assigned]
EPSS is a daily estimate of the probability of exploitation activity being observed over the next 30 days. Following chart shows the EPSS score history of the vulnerability.
CWE - Common Weakness Enumeration

While CVE identifies specific instances of vulnerabilities, CWE categorizes the common flaws or weaknesses that can lead to vulnerabilities. CVE-2017-9657 is associated with the following CWEs:

Common Attack Pattern Enumeration and Classification (CAPEC)

Common Attack Pattern Enumeration and Classification (CAPEC) stores attack patterns, which are descriptions of the common attributes and approaches employed by adversaries to exploit the CVE-2017-9657 weaknesses.

Exploit Prediction

EPSS is a daily estimate of the probability of exploitation activity being observed over the next 30 days.

0.12 }} 0.01%

score

0.45335

percentile

CVSS30 - Vulnerability Scoring System
Attack Vector
Attack Complexity
Privileges Required
User Interaction
Scope
Confidentiality
Integrity
Availability